OMRON Image Processing System FZ5 Series FLV-EP[]Spot Lighting Instruction Sheet
Summary
This comprehensive guide provides essential installation guidelines and safety precautions for Omron's modular lighting system components. Designed for professional electrical engineers, maintenance technicians, and system integrators, the manual ensures safe setup by detailing proper wiring techniques, complying with operational environmental conditions (temperature/humidity), and adhering to critical safety measures. Before use, familiarize yourself with these instructions to guarantee optimal performance, longevity, and compliance of your advanced lighting solution.
Page 1 Text Content
■Specifications
PRECAUTIONS FOR SAFE USE
Model FLV-EP0803W
Please observe the following precautions for safe use of the products.
1. Installation Environment FLV-EP0803R FLV-EP50W
• Do not use the product in environments where it can be exposed to
Item FLV-EP0803B FLV-EP50R
inflammable/explosive gas. • To secure the safety of operation and maintenance, do not install the product FLV-EP0803G close to high-voltage devices and power devices.
• Tighten the screws with the specified torque in this document. FLV-TCC3HB□/-TCC1EP□、FLV-ATC□0405-□
Model FLV-EP□ 2. Power Supply and Wiring
Applicable Controllers W:White、R:Red(Typ.625nm)、B:Blue(Typ.460nm)、G:Green(Typ.525nm)
• For details on power supply and wiring, refer to PRECAUTIONS FOR SAFE
USE on the INSTRUCTION SHEET of the dedicated controller Lighting color (peak wavelength) LED
(FLV-TCC3HB□/-TCC1EP□、FLV-ATC□0405-□).
Light source Risk Group 1(IEC62471)
Spot Lighting • When connecting or disconnecting the lighting connector, be sure to turn off
the power of the unit. Not doing so may result in malfunction. LED Safety White 1.6W 1.6W
• The temperature may reach 70°C while power is being supplied under
Electricity Red 1.1W 1.1W
thermally neutral environment. Do not touch the product directly. Be sure to
INSTRUCTION SHEET wear heat-resistant gloves.
Blue 1.6W
• High-Voltage lines and power lines must be wired separately from this product.
Wiring them together or placing them in the same duct may cause induction, Green 1.6W
Thank you for selecting OMRON product. This sheet resulting in malfunction or damage.
Insulation Resistance 20 MΩ (250VDC)
primarily describes precautions required in installing and • Take sufficient safety measures such as fail-safe circuit to use the product.
3. Others Ambient Temperature Operating: 0 to +50℃, Storage: -15 to +60℃ (with no icing or condensation)
operating the product. • Do not use in safety circuits for atomic energy or that are critical for human life.
Ambient Humidity Operating and storage: 35% to 85% (with no condensation)
• Do not attempt to disassemble, deform by pressure, incinerate, repair, or
Before operating the product, read the sheet thoroughly
modify this product. Degree of Protection IP20(IEC60529) to acquire sufficient knowledge of the product. For your • When disposing of the product, treat as industrial waste.
• Be sure to connect a dedicated lighting controller (FLV-TCC3HB□/-TCC1EP Vibration Resistance(destructive) 10 to 150 Hz, (0.7mm double amplitude) 80 min. each in X, Y, and Z directions
convenience, keep the sheet at your disposal.
□、FLV-ATC□0405-□).Use of other devices may result in fire, explosion, Shock Resistance (destructive) 150 m/s2 3 times each in 6 directions(up/down, left/right, forward/backward) malfunction or failure.
TRACEABILITY INFORMATION: • If you notice an abnormal condition such as a strange odor, extreme heating of Materials Case:Aluminum Representative in EU: Manufacturer:
the unit, or smoke, immediately stop using the product, turn off the power, and Weight Approx.80g Approx.200g
Omron Europe B.V. Omron Corporation,
Wegalaan 67-69 Shiokoji Horikawa, Shimogyo-ku, consult your dealer.
2132 JD Hoofddorp, Kyoto 600-8530 JAPAN • Do not drop or impose excessive vibration or shock on the product. Accessories Instruction sheet (this sheet)
The Netherlands Ayabe Factory It may cause breakdown of the product. 3-2 Narutani, Nakayama-cho, • Ensure that all components which have locking mechanisms are locked before
Ayabe-shi, Kyoto 623-0105 JAPAN using the product.
• Make sure not to drop the product when unpacking or moving it.
The following notice applies only to products that carry the CE mark: 4. Applicable standards
■Connection with lighting controller
Notice: This lighting complied with the EN standard (EN61326-1) This is a class A product. In residential areas it may cause radio (Electromagnetic environment : Industrial electromagnetic interference, in which case the user may be required to take adequate environment (EN/IEC 61326-1 Table 2)) ●Mounting measures to reduce interference. ・ Also, the following condition is applied to the immunity test
of this product. Align the positions of "the tab of the connector at the lighting
There may be cases that Lighting brightness fluctuate Max 10%. cable" and "protrusion of the lighting connector at the lighting
2279062-0E controller", and then insert it until it clicks.
PRECAUTIONS FOR CORRECT USE
© OMRON Corporation 2013 All Rights Reserved.
Observe the following to prevent failure, malfunctioning, and adverse effects Before mounting After mounting on performance and the device. 1. Installation site
Do not install in the following locations: Lighting controller side
• Locations where the ambient temperature exceeds the rated temperature range. PRECAUTIONS ON SAFETY • Locations subject to sudden temperature changes (where condensation will form). • Locations where the relative humidity is below or above 35 to 85% RH. • Locations where there are corrosive or flammable gases. • Locations where there is dust, salt, or iron powder.
● Meanings of Signal Words
• Locations where the device will be subject to direct vibration or shock. Indicates a potentially hazardous situation which, if • Locations where there is strong scattered light (laser light, arc welding light, ultraviolet light, etc.)
not avoided, will result in minor or moderate injury,
WARNING • Locations exposed to direct sunlight or next to a heater. Lighting side
or may result in serious injury or death. Additionally,
• Locations where there is splashing or spraying of water, oil, or chemicals.
there may be significant property damage.
• Locations where there is a strong electrical or magnetic field. 2.Maintenance
Indicates a potentially hazardous situation which, if • Do not install the product close to high-voltage devices and/or power devices
CAUTION not avoided, may result in minor or moderate injury in order to secure the safety of operations and maintenance. ●Removing
or in property damage. • When this product and a cable are attached or removed, be sure to turn OFF
Hold "the tab of the connector at the lighting cable" in the arrow
the power of the Controller first before operation.
• Never use paint thinner, benzen, acetone, or kerosene to clean the product.Wipe (←) direction and pull it straight.
the contamination with a soft cloth included a small amount of alcohol. 【Note】
● Alert Statements in This Sheet
3. Power and cable connections
• When using a commercially available switching regulator, make sure that the Do not pull out with excessive force without pressing the tab.
WARNING Frame ground terminal is grounded.
• If there are surges on your power line, connect a surge absorber as
This product is not designed or rated for ensuring safety appropriate for your conditions of use.
of persons either directly or indirectly. Do not use it for Before removing After removing
• Before turning on the power after the wiring is completed, verify that the
such purposes. power is correct, that there are no incorrect connections such as a shorted
load circuit, and that the load current is suitable. Incorrect wiring may cause Lighting Controller side
Do not disassemble the product. Doing so may cause damage and failures. electric shock due to the high voltage portion. Burn also • Use the product with the specified voltage. Applying a voltage or AC voltage may result due to high temperature. Do not attempt to that exceed the rating may result in burning or explosion of circuit components.
disassemble, deform by pressure, incinerate, repair, or • For cable extension between the lighting and lighting controller, use an modify this product. optional extension cable (FLV-XC□EP/-XC□EPR). Only one extension cable can be used. Looking into the LED light continuously may cause visual • The bending radius at cable wiring must be at least 20 mm.
4.Warming Up Lighting side
impairment. Do not look directly into the LED light.
After turning on the power supply, allow the controller to stand for at least 30 minutes before use. As the circuits are unstable immediately after power on, brightness may gradually change. 5. Lighting electricity
CAUTION
Lighting electricity for lighting equipment to be connected to the Lighting Controller is limited. Be sure to check the lighting electricity written on the
Do not use it exceeding the rated voltage.
plate of the product, label of the packaging box or this manual to connect a
There is a possibility of failure and fire.
lighting within the specified value. 6. Static Electricity
Do not connect amplifier units to AC power supply.
LEDs are easily damaged by static electricity. A grounded wrist strap should
Risk of explosion. thus be used when installing or conducting maintenance on the Diffusion Board.
7. Mounting to Telecentric Lens
Beware of danger of burns.
When using the product mounting to the telecentric lens, do not tighten the
Do not touch the case while the Lighting is ON or immediately
lighting mounting screw too much. Doing so may reduce the brightness due to
after the power is turned OFF, since it is extremely hot.
the stress on the product.
Page Summary Contents For OMRON Image Processing System FZ5 Series FLV-EP[]Spot Lighting Instruction Sheet
Manual Details
| Brand | Image |
|---|---|
| Pages | 2 |
| File Size | 328.73 KB |
| Published | June 05, 2026 |
Enter the captcha to get the download link:
Frequently Asked Questions
What steps should I take if I detect an abnormal condition?
Immediately stop using the product, turn off the power, and consult your dealer if you notice strange odors or extreme heating.
How must high-voltage lines be connected to this device?
High-Voltage lines and power lines must be wired separately from the product to prevent malfunction or damage.
What are the acceptable operating temperature ranges?
The operating range is 0 to +50℃, while the storage range is -15 to +60℃ (without icing).
Is there a specified torque for mounting screws?
Yes, the required tightening torque for securing components is 0.15N·m.