# A B C D E F G H I J K L M N O P Q R S T U V W X Y Z

HANNA HI 250 HI 251 Microprocessor-Based pH mV ºC Bench Meters Instruction Manual

Summary

These microprocessor-based bench meters are essential laboratory instruments designed for logging precise pH, mV, and temperature readings. The comprehensive manual is tailored for technical professionals and lab users needing detailed guidance on setup, operation, calibration protocols, and advanced troubleshooting. It covers everything from initial electrode conditioning to complex data logging functions, ensuring reliable analytical measurement in the field or lab environment.

📄 Preview PAGE OF 40

Page 1 Text Content

Instruction Manual

HI 250 HI 251 Microprocessor-Based p H/m V/ºC Bench Meters

www.hannains com

Page Summary Contents For HANNA HI 250 HI 251 Microprocessor-Based pH mV ºC Bench Meters Instruction Manual

Page 1 Instruction Manual HI 250 HI 251 Microprocessor-Based p H/m V/ºC Bench Meters www.hannains com
Page 2 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary informati...
Page 3 PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or...
Page 4 FUNCTIONAL DESCRIPTION Primary LCD Front Panel Secondary LCD CAL LOG m V p H TIME% DATE INTV BUFFER p H CFM READY BUFWRONG Functions p H Calibration - 3 points GLP Show GLP Data Start/Exit Calibration...
Page 5 HI 250 SPECIFICATIONS Hp0.61ot0.2– Hp00.61ot00.2– EGNAR Cº0.021ot9.9– Hp1.0 Hp10.0 NOITULOSER Hp1.0± Hp10.0± YCARUCCA F°86/C°02@ Cº4.0± )rorreeborpgnidulcxe( ecafretniretupmo C 232SRdetalosi-otpo sref...
Page 6 HI 251 SPECIFICATIONS Hp0.61ot0.2– Hp00.61ot00.2– EGNAR Vm9.996± Vm0002± Cº0.021ot9.9– Hp1.0 Hp10.0 NOITULOSER )Vm9.996±(Vm1.0 )Vm0002±(Vm1 Cº1.0 Hp1.0± Hp10.0± YCARUCCA )Vm9.996±(Vm2.0± F°86/C°02@ )V...
Page 7 OPERATIONAL GUIDE POWER CONNECTION Plug the 12 VDC adapter into the power supply socket. Note: These instruments use non volatile memory to retain the p H, m V, temperature calibrations and all other ...
Page 8 If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample to prevent cros...
Page 9 RELATIVE m V MEASUREMENTS (HI 251 only) Press RANGE until “r EL” message will be displayed on the secondary LCD line for about one second and “m V” tag will blink. After one second the temperature wil...
Page 10 p H CALIBRATION Calibrate the instrument frequently, especially if high accuracy is required. The instrument should be recalibrated : Whenever the p H electrode is replaced. At least once a week. Afte...
Page 11 The “ ” tag will blink on the LCD until the reading is stable. When the reading is stable and close to the selected buffer, “READY” tag will be displayed and “CFM” tag will blink. Press CFM to confirm...
Page 12 Note: If the value measured by the instruments is not close to the selected buffer, “WRONG” “ BUF ” and “WRONG” “ ” tags will blink alternately. In this case, check if the correct buffer has been used...
Page 13 GOOD LABORATORY PRACTICE (GLP) GLP is a set of functions that allows storage and retrieval of data regarding the maintenance and status of the electrode. All data regarding last calibration (one, two ...
Page 14 The calibration buffers in calibrating order. The first p H calibration buffer: CAL p H The second p H calibration buffer: CAL p H The third p H calibration buffer: CAL p H Note: The “OLd” message dis...
Page 15 The instrument ID. Note: • Press GLP at any moment and the instruments will return to measurement mode. If calibration has not been performed, the instruments display “no CAL” message blinking. CAL LO...
Page 16 VIEW LOGGED DATA Press RCL while in measurement mode to retrive the stored information. If no data were logged, the instruments display: Otherwise, the instruments will display the p H or relative m V...
Page 17 • The date on the primary LCD, along with “DATE” tag. LOG DATE The calibration offset on the primary LCD and “OFS” message on the secondary LCD. LOG m V The calibration slope on the primary LCD and “S...
Page 18 SETUP Setup mode allows viewing and modifying the following parameters: Current Time Current Date p H Resolution Beep Status Baud Rate (serial communication) Command prefix (serial communication) Cali...
Page 19 CURRENT DATE SET Press CAL when the current date is displayed. The year will start blinking. Press the ARROW keys to change the year. Press RANGE. The month will start blinking. Press the ARROW keys t...
Page 20 Press the ARROW keys to change the beep status (On or OFF). Press CFM to save the modified beep status. Press CAL to escape without saving. BAUD RATE SET Press CAL when the baud rate is displayed. The...
Page 21 INSTRUMENT ID PARAMETER SET Press CAL when the instrument ID is displayed. The instrument ID (0000 to 9999) will start blinking. Press the ARROW keys to change the instrument ID value. Press CFM to sa...
Page 22 TEMPERATURE CALIBRATION(for technical personnel only) All the instruments are factory calibrated for temperature. Hanna’s temperature probes are interchangeable and no temperature calibration is neede...
Page 23 Use the ARROW keys to set the reading on the secondary LCD to that of the hot water. When the reading is stable and close to the selected calibration point, “READY” tag will appear and “CFM” tag will ...
Page 24 Note: If the reading is not close to the selected calibration point, “WRONG” tag will blink. Verify calibration condition or contact your vendor if you can not calibrate. Press CAL in any moment of th...
Page 25 PC INTERFACE Data transmission from the instrument to the PC can be done with the HI 92000 Windows® compatible software (optional). HI 92000 also offers graphing and on-line help feature. Data can be ...
Page 26 CHR n Change the instrument range according with the parameter value (n): n=1 p H range/0.01 resolution n=2 p H range/0.1 resolution n=3 m V range (HI 251 only) n=4 Relative m V range (HI 251 only) CO...
Page 27 p H VALUES AT DIFFERENT TEMPERATURES The temperature has an effect on p H. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibra...
Page 28 ELECTRODE CONDITIONING & MAINTENANCE PREPARATION PROCEDURE Remove the protective cap of the p H electrode. DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will...
Page 29 For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction or HI 7071 or HI ...
Page 30 CLEANING PROCEDURE • General Soak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately ½ hour. • Protein Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Solution for 15 minutes. •...
Page 31 TROUBLESHOOTING GUIDE SMOTPMYS MELBORP NOITULOS evissecxe/esnoperwol S .edortcele Hpytri D nipitedortceleehtkao S .tfird 1608IHro1607IH setunim03rofnoitulos ehtnaelcnehtdna .edortcele puetautculfsgnid...
Page 32 TEMPERATURE CORRELATION FOR p H SENSITIVE GLASS The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes more time for th...
Page 33 ACCESSORIES p H BUFFER SOLUTIONS HI 70004P p H 4.01 Buffer Sachets, 20 m L, 25 pcs HI 70007P p H 7.01 Buffer Sachets, 20 m L, 25 pcs HI 70010P p H 10.01 Buffer Sachets, 20 m L, 25 pcs HI 7001L p H 1.6...
Page 34 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution, 460 m L HI 7092L Oxidizing Pretreatment Solution, 460 m L p H ELECTRODES All electrodes part numbers ending in B are supplied with a...
Page 35 HI 1330B Glass-body, semimicro, single junction, refillable, combination p H electrode. Use: laboratory, vials. 5mm DIA 5mm 0.2" 0.2" mm 4.7" "S" VERSION HI 1331B Glass-body, ...
Page 36 FC 100B Plastic-body (Kynar®), double junction, refillable, combination p H electrode. Use: general purpose for food industry. mm 0.5" mm 4.7" FC 200B Plastic-body (Kynar®), open junction, c...
Page 37 HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination p H electrode. Use: surface measurement. mm 0.5" mm 4.3" ORP ELECTRODES HI 3131B Glass-body, refillable...
Page 38 EXTENSION CABLE FOR SCREW-TYPE ELECTRODES (SCREW TO BNC ADAPTER) HI 7855/1 Extension cable 1 m (3.3') long HI 7855/3 Extension cable 3 m (9.9') long HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3.0 m...
Page 39 RECOMMENDATIONS FOR USERS Before using these products, make sure they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause ...
Page 40 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 ...

Manual Details

Brand Bench
Pages 40
File Size 699.96 KB
Published May 16, 2026
103 views

Enter the captcha to get the download link:

captcha

Frequently Asked Questions

What type of measurements can the HI 251 model perform?

It is a logging microprocessor-based pH and temperature bench meter that can also measure Oxidation-Reduction Potential (ORP).

Are there safety precautions when handling the electrodes?

Wear ESD wrist straps to avoid damage from electrostatic discharges, and do not touch the glass bulb at all times.

What must be done if I need service or repair?

Contact your dealer. If under warranty, report the model number, date of purchase, serial number, and nature of the problem.

Is documentation included when purchasing the unit?

Yes, each instrument is supplied with an Instruction Manual and essential accessories like electrode and buffer solutions.