Page 1
Instruction Manual 748468-D July 2003 Model Micro CEM SHS Sample Handling System http://www.processanalytic.com
Page 2
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures and tests its products to meet many national and in- ternational standards. Because these instrument...
Page 3
Instruction Manual 748468-D Model Micro CEM SHS July 2003 TABLE OF CONTENTS PREFACE........................................................................................................................
Page 4
Instruction Manual 748468-D July 2003 Model Micro CEM SHS LIST OF ILLUSTRATIONS Figure 1-1. Micro CEM Sample Handling System........................................................................ 1-1...
Page 5
Instruction Manual 748468-D Model Micro CEM SHS July 2003 PREFACE The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Micro C...
Page 6
Instruction Manual 748468-D July 2003 Model Micro CEM SHS INTENDED USE STATEMENT The Micro CEM Sample Handling System is intended for use as an industrial process measurement device only. It is not in...
Page 7
Instruction Manual 748468-D Model Micro CEM SHS July 2003 WARNING. PARTS INTEGRITY AND UPGRADES Tampering with or unauthorized substitution of components may adversely affect the safety of this instru...
Page 8
Instruction Manual 748468-D July 2003 Model Micro CEM SHS GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compressed Gas Association's &...
Page 9
Instruction Manual 748468-D Model Micro CEM SHS July 2003 DOCUMENTATION The following Micro CEM SHS instruction materials are available. Contact Customer Service Center or the local representative to ...
Page 10
Instruction Manual 748468-D July 2003 Model Micro CEM SHS P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Page 11
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 1 DESCRIPTION AND SPECIFICATIONS The sampled gas is collected from the stack 1-1 OVERVIEW and prepared by the Probe/Sample Handling Sy...
Page 12
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 1-2 Micro CEM SHS SPECIFICATIONS Power Universal Power Supply 85 – 264 VAC, 50 – 60 Hz, + 10% 500 Watts Maximum at Start Up. 250 Watts Nominal...
Page 13
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 2 INSTALLATION WARNING 2-1 INSTALLATION ELECTRICAL SHOCK HAZARD Refer to drawings located in the rear of this manual. POSSIBLE EXPLOSI...
Page 14
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 5. Connect analog output signal as required NOTE per Electrical Interconnect and Outline Refer to all NOTES on drawings located in and Mountin...
Page 15
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 3 OPERATION 4. Verify that analyzer ranges and analog CAUTION. outputs are configured as desired. Do not operate or service this instr...
Page 16
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 8. Verify output from analyzer. Adjust as pare this value with zero and span gas ad- required. Refer to analyzer instruction mitted directly t...
Page 17
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 4 MAINTENANCE AND SERVICE The Sample Conditioning system is designed sary, to maintain a high level of data for minimum maintenance ba...
Page 18
Instruction Manual 748468-D July 2003 Model Micro CEM SHS WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. SP1 Primary sintered metal sample ...
Page 19
Instruction Manual 748468-D Model Micro CEM SHS July 2003 WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. Packing Adjust- Regulating/Needle ...
Page 20
Instruction Manual 748468-D July 2003 Model Micro CEM SHS WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. Indication V FI1 Flowmeters O-Ring...
Page 21
Instruction Manual 748468-D Model Micro CEM SHS July 2003 WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. Coolant and/or PR1 Refrigerated Co...
Page 22
Instruction Manual 748468-D July 2003 Model Micro CEM SHS WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. Heated/Vaporizing Pressure Regulat...
Page 23
Instruction Manual 748468-D Model Micro CEM SHS July 2003 WHERE EVERY EVERY EVERY EVERY WEEKLY MONTHLY USED COMPONENT MAINTENANCE ITEM DAILY 3 MO. 6 MO. 9 MO. 12 MO. Purge Pressure Switches T Sample P...
Page 24
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 4-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Page 25
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 5 TROUBLESHOOTING DANGER. ELECTRICAL SHOCK HAZARD Disconnect power to the module(s) prior to replacing components. WARNING QUALIFIED P...
Page 26
Instruction Manual 748468-D July 2003 Model Micro CEM SHS SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION a. Sample Flow valves not adjusted properly a. Adjust correct flow for analyzer(s) b. No sample press...
Page 27
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 6 REPLACEMENT PARTS 6-1 INTERNAL SAMPLE PANEL 6-2 SAMPLE PROBES 662210 Relay Board Assembly 662011 Probe Assembly 5 foot 897861 Fuse 5...
Page 28
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 6-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Page 29
Instruction Manual 748468-D Model Micro CEM SHS July 2003 SECTION 7 RETURN OF MATERIAL If warranty service is expected, the defective 7-1 RETURN OF MATERIAL unit will be carefully inspected and tested...
Page 30
Instruction Manual 748468-D July 2003 Model Micro CEM SHS 7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Page 31
WARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) mon...
Page 32
Instruction Manual 748468-D July 2003 Model Micro CEM SHS EMERSON PROCESS MANAGEMENT Rosemount Analytical Inc. EUROPEAN TECHNOLOGY CENTER Process Analytic Division Fisher-Rosemount Gmb H & Co. 120...