Fluke 17B MAX Digital Multimeter 15B Max 17B Calibration Manual

Summary

This comprehensive technical guide provides authorized service technicians with detailed instructions for maintaining and calibrating 15B MAX and 17B MAX digital multimeters. The manual covers critical procedures, including general maintenance checks, fuse testing, and professional adjustment techniques using specialized equipment. It directs users on entering calibration modes, making precise electrical connections, and executing voltage, current, and temperature calibrations to ensure the meter operates with maximum accuracy. This essential resource is designed for skilled repair personnel needing expert-level servicing of Fluke multimeters.

📄 Preview 📖 Table of Contents Contents PAGE OF 14

Page 1 Text Content

15B MAX/17B MAX Digital Multimeters

Calibration Manual

August 2022 (English) ©2022 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

Page Summary Contents For Fluke 17B MAX Digital Multimeter 15B Max 17B Calibration Manual

Page 1 15B MAX/17B MAX Digital Multimeters Calibration Manual August 2022 (English) ©2022 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are tr...
Page 2 LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is 1 year and begins ...
Page 3 Table of Contents Introduction................................................................................................................................... Contact Fluke Safety Information.........
Page 4 15B MAX/17B MAX Calibration Manual
Page 5 Introduction The 15B MAX/17B MAX Digital Multimeters (the Product) are 6000 count instruments. The Product is battery powered with a digital display. Unless otherwise identified, all illustrations sho...
Page 6 15B MAX/17B MAX Calibration Manual XW Warning To prevent possible electrical shock, fire, or personal injury, and for safe operation and maintenance of the product, see Safety Information. General Mai...
Page 7 Digital Multimeters Maintenance Static Awareness Semiconductors and integrated circuits can be damaged by electrostatic discharge during handling. This notice explains how to minimize damage to these ...
Page 8 15B MAX/17B MAX Calibration Manual Replace Batteries and Fuses XW Warning To prevent false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as th...
Page 9 Digital Multimeters Performance Tests Table 1 lists the equipment required for the performance tests. Table 1. Required Equipment Equipment Recommended Model Calibrator Fluke Calibration 5502A, 5502E,...
Page 10 15B MAX/17B MAX Calibration Manual Table 2. Performance Specifications (cont.) Calibrator Output 15B 17B Lower Upper Step Function Frequency or MAX MAX Limit Limit Value Amplitude BEEPER ON 40 Ω DIODE...
Page 11 Digital Multimeters Adjustment Procedures Figure 3. Temperature Connections 5522A CALIBRATOR STBY OPR EARTH SCOPEEXGRD MENU d Bm sec EDIT Hz SETUP RESET FIELD Fn NEW CA REF CE MEAS MORE TC MODES SHIFT...
Page 12 15B MAX/17B MAX Calibration Manual Figure 4. Remove Calibration Sticker 3. See Figure 5 to make the connections to power and the calibrator. Use a contact spring to connect a 3 V dc power supply. Conn...
Page 13 Digital Multimeters Adjustment Procedures 4. To put the DUT in adjustment mode, use a small probe to push the calibration button (CAL1). See Figure 6. Figure 6. Adjustment WP7 CAL1 WP8 WP6 WP9 HOLD1 5...
Page 14 15B MAX/17B MAX Calibration Manual 8. When you complete adjustment, turn off the DUT. Table 5. All Function Adjustments Function Step Display Calibrator Output MAX MAX VAC High voltage mark (Y) flicke...

Manual Details

Brand Fluke
Pages 14
File Size 1.02 MB
Published June 05, 2026
43 views

Enter the captcha to get the download link:

captcha

Frequently Asked Questions

How do I enter the device for calibration mode?

Remove the battery door and batteries, remove the calibration sticker, and then use a small probe to push the specific calibration button (CAL1).

What equipment is required to perform manual adjustments on the 17B MAX?

You need an adjustable potentiometer (Type K) connected via a Fluke 80AK-A adapter to dual banana jacks.

How do I adjust the temperature function on the 17B MAX?

Use an adjustable potentiometer for manual adjustment, ensuring it reads between 0.1 ÁC and 0.3 ÁC when the input is at 0 °C.