This specialized adapter facilitates the seamless integration of Varian CT-100 gauges into vacuum systems originally equipped with Edwards ATC or APG gauges. It allows technicians to install modern vacuum technology without the hassle of removing existing cabling connected via RJ45 connectors. This guide details the adapter's precise wiring, electrical compatibility, and implementation process, making it essential for engineers and service professionals managing complex mixed-vendor vacuum infrastructure.

📄 Preview 📖 Table of Contents Contents PAGE OF 4

Page 1 Text Content

vacuum technologies

FIELD INSTRUCTION SHEET

Edwards ATC and APG to Varian CT-100 Pin Adapter

Manual No. 699908071 Revision A Date August 2002

Page Summary Contents For Agilent Technologies Edwards ATC and APG to Varian CT-100 Pin Adapter Field Instruction Sheet

Page 1 vacuum technologies FIELD INSTRUCTION SHEET Edwards ATC and APG to Varian CT-100 Pin Adapter Manual No. 699908071 Revision A Date August 2002
Page 2 This page intentionally left blank.
Page 3 Field Instruction Sheet Varian Vacuum Technologies Edwards ATC & APG to Varian CT-100 Pin Adapter The L7439301 Adapter enables the installation of Varian’s CT-100 Active Rough Vacuum Gauge into sy...
Page 4 Sales and Service Offices Argentina Germany and Austria UK and Ireland Varian Argentina Ltd. Varian Deutschland Gmb H Varian Ltd. Sucursal Argentina Alsfelder Strasse 6 28 Manor Road Av. Ricardo Balbi...

Manual Details

Brand Agilent Technologies
Pages 4
File Size 40.96 KB
Published May 13, 2026
83 views

Enter the captcha to get the download link:

captcha

Frequently Asked Questions

What is the purpose of the L7439301 Adapter?

It enables Varian’s CT-100 Active Rough Vacuum Gauge to be installed with existing Edwards ATC or APG systems without removing the current RJ45 cable.

How do I install the adapter?

First, plug the existing system RJ45 connector into the Adapter, and then follow the CT-100 manual for installation.

Will the CT-100 signal match the Edwards gauge output?

No. The analog output of the CT-100 differs significantly from Edwards gauges; adjustments are necessary for accurate display unless using setpoint signals.