This essential service manual provides comprehensive instructions for replacing the Ion Chamber in the VS C15 Leak Detector. Designed specifically for operators and field service personnel, this guide covers best practices, critical safety warnings, fluid handling, and proper cleaning techniques necessary for maintaining accurate leak detection performance. By thoroughly reviewing these procedures, users can ensure the correct installation and reliable operation of their vacuum equipment.

📄 Preview 📖 Table of Contents Contents PAGE OF 16

Page 1 Text Content

vacuum technologies

FIELD INSTALLATION

Ion Chamber

INSTRUCTIONS

Replacement for VS C15 Leak Detector

Part Number 699910015 Rev. A December 2008

Page Summary Contents For Agilent Ion Chamber Replacement for VS C15 Leak Detector Owner's Manual

Page 1 vacuum technologies FIELD INSTALLATION Ion Chamber INSTRUCTIONS Replacement for VS C15 Leak Detector Part Number 699910015 Rev. A December 2008
Page 2 Ion Chamber for VS C15 Leak Detector /1 /0 This page intentionally left blank.
Page 3 Varian Field Instruction Sheet Ion Chamber Replacement for VS C15 Leak Detector Preface Documentation Standards This manual uses the following documentation standards: NOTE Notes contain important inf...
Page 4 Operators and service personnel must be aware of all hazards associated with this equipment. They must know how to recognize hazardous and potentially hazardous conditions, and know how to avoid them....
Page 5 O-ring Care When removing, checking or replacing O-rings, keep in mind the following: NOTE Varian recommends replacing all O-rings during routine maintenance or during any maintenance procedure requir...
Page 6 Equipment Required Metric Allen Wrench (Range 1.5mm to 10mm) M3 Philips head screwdriver Non-powdered ESD-safe Nitrile or equivalent gloves Isopropyl alcohol Torque wrench, M5 head (must be adjustable...
Page 7 4. If the unit was just running, either: Open the turbo vent valve, if the internal calibrated leak option was not installed (Figure 2: Turbo Vent), Or crack the NW16 connection (Figure 3: NW-16 Vent)...
Page 8 5. Disconnect both cables from the spectrometer (Figure 4: Cables To Spectrometer). Figure 4: Cables To Spectrometer CAUTION Static sensitive device, ensure that personnel are properly grounded before...
Page 9 7. Using an M5 allen wrench, remove the six screws holding on the spectrometer header assembly (Figure 6: Spec- trometer Header Assembly Screws). Figure 6: Spectrometer Header Assembly Screws NOTE Wea...
Page 10 9. Remove the O-ring, inspect and replace if damaged (Figure 8: O-ring). Figure 8: O-ring 10.If the inside of the spectrometer body appears to be dirty, clean it with a lint free wipe and Isopropyl al...
Page 11 CAUTION This is a static sensitive device, ensure that personnel are properly grounded before proceeding. 11.Use an M1.5 Allen wrench to unscrew the six coupler screws and the collimator screw (Figure...
Page 12 CAUTION Ensure that no damage is done to the spectrometer wire during disassembly. 15.Loosen the two screws on the rear and slide the ion chamber out (Figure 12: Partial Pressure Plate). Ceramic Remov...
Page 13 Figure 13: Extractor Plate 21.Reinstall the header into the body with the detector housing closest to the inlet with the six existing screws. Torque to 10.2 N-M (90 IN-LB). NOTE The cones of the Belle...
Page 14 24.Re-plug in both cables and ground strap. 25.Verify that turbo vent valve is closed. 26.Leak check the spectrometer to ensure a leak free joint. 27.Close the cover and secure using existing hardware...
Page 16 Sales and Service Offices Canada Italy United States Central coordination through: Varian, Inc. Varian, Inc Varian, Inc. via F.lli Varian 54 121 Hartwell Avenue 121 Hartwell Avenue 10040 Leini, (Torin...

Manual Details

Brand ION
Pages 16
File Size 941.69 KB
Published May 24, 2026
117 views

Enter the captcha to get the download link:

captcha

Frequently Asked Questions

How should I clean the leak detector enclosure and display?

Use only a soft cloth slightly dampened with water or a mild soap; do NOT use excess water or strong solvents.

What is recommended for cleaning vacuum system components?

VacuSolv solvent, Varian Component and Spectrometer Cleaning Kit, IPA, or ® Dow Corning OS-20 are acceptable alternatives.

When should O-rings be replaced during maintenance?

Varian recommends replacing all O-rings during routine maintenance or any procedure requiring their removal to ensure a proper seal.